# Checklist: what a cyber device’s premarket submission needs under section 524B and FDA’s February 2026 guidance

Thirteen steps for medical device makers assembling the cybersecurity content of a 510(k), PMA, De Novo, PDP or HDE, separating what section 524B requires from what FDA’s February 3, 2026 guidance recommends.

Checklist · Healthcare · Platforms & Architecture. Published November 14, 2026. Last reviewed November 14, 2026.

## Statute and guidance

Section 524B of the Federal Food, Drug, and Cosmetic Act, codified at 21 U.S.C. 360n-2, was added by section 3305 of the Food and Drug Omnibus Reform Act of 2022 (Public Law 117-328), enacted December 29, 2022, and took effect 90 days later [1]. It requires anyone submitting a 510(k), PMA, PDP, De Novo or HDE for a “cyber device” to include the information FDA requires to show the device meets four cybersecurity requirements [1]. FDA’s guidance “Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions,” issued February 3, 2026, supersedes the June 27, 2025 version [2] [3]. It aligns the recommendations with the Quality Management System Regulation in 21 CFR part 820, which took effect on February 2, 2026 and incorporates ISO 13485:2016 by reference [3].

## Reading the steps

FDA says its guidance should be read as recommendations unless specific statutory or regulatory requirements are cited [3]. Each step below says which it is: a requirement in section 524B, or an FDA recommendation from a named section of the guidance. It is written for regulatory affairs, product security and quality teams. The owner under each step is a typical assignment; the statute places the duties on the person who submits and on the sponsor [1]. Not legal advice. FDA recommends its Q-Submission process for discussing cybersecurity design considerations with the agency [3].

## The checklist

1. **Decide, and document why, whether the device is a cyber device: sponsor-authorized software, the ability to connect to the internet, and characteristics that could be vulnerable to cybersecurity threats.** (Regulatory affairs with product security) Section 524B(c) defines a cyber device by those three criteria [1]. FDA reads them broadly: software includes firmware and programmable logic, and the ability to connect includes Wi-Fi, cellular, Bluetooth, magnetic inductive links and hardware connectors such as USB, ethernet or serial ports, whether or not connection was intended (guidance section VII.B) [3].
2. **Confirm the submission type triggers section 524B, including 510(k)s and PMA supplements for modifications of a cyber device.** (Regulatory affairs) Section 524B(a) applies to submissions under the 510(k), PMA, PDP, De Novo and HDE pathways [1]. FDA reads this to cover original, special and abbreviated 510(k)s, PMA and HDE supplements, and device modifications that need one of those submissions (guidance sections VII.A and VII.D) [3].
3. **Write the postmarket cybersecurity plan to monitor, identify and address vulnerabilities and exploits, including coordinated vulnerability disclosure.** (Product security incident response team) Required by section 524B(b)(1) [1]. FDA recommends the plan name responsible personnel, sources and frequency of vulnerability monitoring, handling of CISA Known Exploited Vulnerabilities, periodic testing, patch timelines and update processes, the disclosure process, and how remediations will be communicated to customers (guidance sections VI.B and VII.C.1) [3].
4. **Set and justify a regular update cycle for known unacceptable vulnerabilities, and a process for out-of-cycle patches for critical ones.** (Engineering and product security) Section 524B(b)(2) requires updates and patches on a reasonably justified regular cycle for known unacceptable vulnerabilities, and as soon as possible out of cycle for critical vulnerabilities that could cause uncontrolled risks [1]. FDA recommends the justification for the cycle go in the plan and reflect the device’s risk in its environment of use (guidance section VII.C.1) [3].
5. **Show the design and maintenance processes that give reasonable assurance the device and related systems are cybersecure, ideally as a secure product development framework within the quality system.** (Quality and engineering leadership) Section 524B(b)(2) requires such processes and procedures [1]. FDA counts update servers, other manufacturer-controlled devices and connections to healthcare facility networks among related systems, and describes a secure product development framework as one way to satisfy the QMSR, not the only way (guidance sections IV.A.1 and VII.C.2) [3].
6. **Prepare the threat model and cybersecurity risk assessment, kept current across the device’s lifecycle.** (Product security) Recommended for all premarket submissions for devices with cybersecurity risk (guidance sections V.A.1, V.A.2 and Appendix 4) [3]. FDA also recommends updating these documents as new threats and vulnerabilities are found, which can help satisfy the patching requirements of section 524B(b)(2) (guidance section VII.C.1) [3].
7. **Produce a machine-readable SBOM covering commercial, open-source and off-the-shelf components, with each component’s support level and end-of-support date.** (Software engineering) An SBOM is required by section 524B(b)(3) [1]. FDA recommends the NTIA minimum elements plus, for each component, the level of support and the end-of-support date, in the SBOM or an addendum (guidance sections V.A.4.b and VII.C.3) [3].
8. **List every known vulnerability in the device and its components, including any in CISA’s Known Exploited Vulnerabilities Catalog, with a risk assessment and controls for each.** (Product security) Recommended by FDA, which also asks how each vulnerability was found so it can judge whether the assessment methods were adequate (guidance section V.A.4.b) [3].
9. **Assess the security impact of the software anomalies that remain unresolved at the time of submission.** (Software quality assurance) Recommended by FDA, because some anomalies may also be vulnerabilities (guidance section V.A.5) [3].
10. **Draw the security architecture views: global system, multi-patient harm, updateability and patchability, and security use case views.** (Systems architecture) Recommended by FDA, with additional views where the system is too complex for a single global system view (guidance section V.B and Appendices 2 and 4) [3].
11. **Submit the security testing evidence: requirements testing, threat mitigation testing, vulnerability testing and penetration testing, with the testers’ independence stated.** (Security verification and validation) Recommended by FDA, including fuzz testing, attack surface analysis, software composition analysis and static and dynamic code analysis; penetration test reports should give scope, duration, methods, results and the testers’ independence, with original third-party reports attached (guidance section V.C) [3].
12. **Draft the security sections of the labeling: recommended controls for the use environment, a list of network ports and interfaces, and infrastructure requirements.** (Labeling and technical writing, with product security) Recommended by FDA as a way to meet labeling requirements under sections 502(a)(1) and 502(f) of the FD&C Act for devices with cybersecurity risk (guidance section VI.A) [3].
13. **For a modification, classify the change: one that may affect cybersecurity needs the full documentation; one unlikely to still needs the section 524B(b)(1) plan if none was filed before.** (Regulatory affairs) FDA gives changes to authentication or encryption, new connectivity and changes to the software update mechanism as examples that may impact cybersecurity, and changes to materials or sterilization as unlikely to (guidance section VII.D) [3]. Section 524B(b)(4) also lets FDA add requirements by regulation [1]; check for any before filing.

## Sources

1. United States Code, 2024 edition (via GovInfo). 21 U.S.C. 360n-2, Ensuring cybersecurity of devices (FD&C Act section 524B). <https://www.govinfo.gov/content/pkg/USCODE-2024-title21/html/USCODE-2024-title21-chap9-subchapV-partA-sec360n-2.htm> Checked November 14, 2026.
2. U.S. Food and Drug Administration. Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions (guidance page). <https://www.fda.gov/regulatory-information/search-fda-guidance-documents/cybersecurity-medical-devices-quality-management-system-considerations-and-content-premarket> Checked November 14, 2026.
3. U.S. Food and Drug Administration. Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, final guidance issued February 3, 2026 (PDF). <https://www.fda.gov/media/119933/download> Checked November 14, 2026.